Resolving fluorescence beads at 100-200 nm axial distance with a two-photon 4Pi-microscope operating in the near infrared




Hell S., Schrader M., Hänninen P., Soini E.

1995

 Optics Communications

Optics Communications

120

null

129

133

5

0030-4018

http://api.elsevier.com/content/abstract/scopus_id:0029388396



We show the possibility of resolving point objects having an axial distance of 100-200 nm with a precision of ± 15 nm with a two-photon excitation 4Pi-microscope. The resolution is demonstrated by axial imaging of clustered fluorescence beads. A comparison of the axial resolution of the two-photon excitation 4Pi-microscope with a regular two-photon excitation microscope shows a four-fold improvement in axial resolution. © 1995.



Last updated on 13/10/2025 12:29:51 PM