A1 Refereed original research article in a scientific journal

Resolving fluorescence beads at 100-200 nm axial distance with a two-photon 4Pi-microscope operating in the near infrared




AuthorsHell S., Schrader M., Hänninen P., Soini E.

Publication year1995

Journal:Optics Communications

Journal name in sourceOptics Communications

Volume120

Issuenull

First page 129

Last page133

Number of pages5

ISSN0030-4018

Web address http://api.elsevier.com/content/abstract/scopus_id:0029388396


Abstract
We show the possibility of resolving point objects having an axial distance of 100-200 nm with a precision of ± 15 nm with a two-photon excitation 4Pi-microscope. The resolution is demonstrated by axial imaging of clustered fluorescence beads. A comparison of the axial resolution of the two-photon excitation 4Pi-microscope with a regular two-photon excitation microscope shows a four-fold improvement in axial resolution. © 1995.