OPTIMIZATION OF H-2 O-15-CBF METHOD FOR HIGHER SIGNAL-TO-NOISE RATIO




KANNO I, IIDA H, MIURA S

PublisherKLUWER ACADEMIC PUBL

1991

Medical progress through technology

MEDICAL PROGRESS THROUGH TECHNOLOGY

MED PROG TECHNOL

17

3-4

189

192

4

0047-6552



We described theoretical principle and data analysis for PET study using intravenous (H2O)-O-15 for measurement of cerebral blood flow (CBF) with emphasis on stimulation study for psychophysiological test. In order to improve signal-to-noise (S/N) ratio in CBF images, we optimized the scan procedures and data analysis. We determined optimal scantime of autoradiographic image to provide minimum S/N ratio as to 90 through 120 sec. Increase of full-width at half-maximum (FWHM) of smoothing filter did reduce the image noise (standard deviation, SD) more rapidly than the partial volume effect. Phantom experiment showed a tendency that a larger FWHM provided a higher statistical t-value.