TekijätNesterov Sergey V., Sciagrà Roberto, Orozco Luis Eduardo Juarez, Prior John O., Settimo Leonardo, Han Chunlei, Deshayes Emmanuel, deKemp Robert A., Ryzhkova Darja V., Gwet Kilem L., Knuuti Juhani M.

KustantajaSpringer

Julkaisuvuosi2022

JournalJournal of Nuclear Cardiology

Tietokannassa oleva lehden nimiJOURNAL OF NUCLEAR CARDIOLOGY

Lehden akronyymiJ NUCL CARDIOL

Vuosikerta29

Aloitussivu2543

Lopetussivu2550

Sivujen määrä8

ISSN1071-3581

eISSN1532-6551

DOIhttps://doi.org/10.1007/s12350-021-02741-4

Verkko-osoitehttps://link.springer.com/article/10.1007/s12350-021-02741-4

Rinnakkaistallenteen osoitehttps://research.utu.fi/converis/portal/detail/Publication/66898194


Tiivistelmä

Purpose

To cross-compare three software packages (SPs)-Carimas, FlowQuant, and PMOD-to quantify myocardial perfusion at global, regional, and segmental levels.

Materials and Methods

Stress N-13 ammonia PET scans of 48 patients with HCM were analyzed in three centers using Carimas, FlowQuant, and PMOD. Values agreed if they had an ICC > 0.75 and a difference < 20% of the median across all observers.

Results

When using 1TCM on the global level, the agreement was good, and the maximum difference between 1TCM MBF values was 17.2% (ICC = 0.83). On the regional level, the agreement was acceptable except in the LCx region (25.5% difference, ICC = 0.74) between FlowQuant and PMOD. Carimas-1TCM agreed well with PMOD-1TCM and FlowQuant-1TCM. Values obtained with FlowQuant-1TCM had a somewhat lesser agreement with PMOD-1TCM, especially at the segmental level.

Conclusions

The global and regional MBF values (with one exception) agree well between the different software packages. There is significant variability in segmental values, mainly located in the LCx region and segments. Out of the studied tools, Carimas can be used interchangeably with both PMOD and FlowQuant for 1TCM implementation on all levels-global, regional, and segmental.