A Respiratory Motion Estimation Method Based on Inertial Measurement Units for Gated Positron Emission Tomography




Lehtonen Eero, Teuho Jarmo, Koskinen Juho, Jafari Tadi Mojtaba, Klén Riku, Siekkinen Reetta, Gambin Joaquin R., Vasankari Tuija, Saraste Antti

PublisherMDPI

2021

Sensors

SENSORS

SENSORS-BASEL

ARTN 3983

21

12

22

DOIhttps://doi.org/10.3390/s21123983

https://research.utu.fi/converis/portal/detail/Publication/66434172



We present a novel method for estimating respiratory motion using inertial measurement units (IMUs) based on microelectromechanical systems (MEMS) technology. As an application of the method we consider the amplitude gating of positron emission tomography (PET) imaging, and compare the method against a clinically used respiration motion estimation technique. The presented method can be used to detect respiratory cycles and estimate their lengths with state-of-the-art accuracy when compared to other IMU-based methods, and is the first based on commercial MEMS devices, which can estimate quantitatively both the magnitude and the phase of respiratory motion from the abdomen and chest regions. For the considered test group consisting of eight subjects with acute myocardial infarction, our method achieved the absolute breathing rate error per minute of 0.44 +/- 0.23 1/min, and the absolute amplitude error of 0.24 +/- 0.09 cm, when compared to the clinically used respiratory motion estimation technique. The presented method could be used to simplify the logistics related to respiratory motion estimation in PET imaging studies, and also to enable multi-position motion measurements for advanced organ motion estimation.