Quantitative Detection of Naproxen Surface Residue Using FTIR Spectroscopy and Partial Least Squares Regression;




AuthorsFaraz, Mehdi; Heikkonen, Jukka; Ranti, Tuomas; Savola, Anna; Heikkinen, Susanna; Kiljunen, Eero; Mäkilä, Tuomas

PublisherInstitute of Electrical and Electronics Engineers (IEEE)

Publication year2026

Article number4502004

Volume10

Issue7

eISSN2475-1472

DOIhttps://doi.org/10.1109/LSENS.2026.3697991

Publication's open availability at the time of reportingOpen Access

Publication channel's open availability Partially Open Access publication channel

Web address https://doi.org/10.1109/lsens.2026.3697991

Self-archived copy’s web addresshttps://research.utu.fi/converis/portal/detail/Publication/527055284

Self-archived copy's licenceCC BY

Self-archived copy's versionPublisher`s PDF


Abstract

Cleaning validation is essential in pharmaceutical manufacturing to prevent cross-contamination, yet traditional wipe-sampling and chromatographic methods are labor-intensive and time-consuming. This study demonstrates handheld Fourier transform infrared (FTIR) spectroscopy combined with partial least squares (PLS) regression as a rapid, nondestructive alternative for quantifying naproxen surface residue on stainless steel. A total of 293 FTIR spectra were collected from stainless steel coupons spiked with naproxen (0 to 5.15 μg cm−2) over 650 cm−1–4000 cm−1. After outlier removal (IQR method), 263 spectra were retained and preprocessed using Savitzky–Golay smoothing and standard normal variate (SNV) normalization. The global PLS model (fivefold CV) achieved R2=0.854, RMSE = 0.491 μg cm−2, limit of detection (LOD) = 1.95 μg cm−2, and limit of quantification (LOQ)= 5.92 μg cm−2. Interval PLS identified the 1646–1810 cm−1 region (naproxen C = O stretch) as most predictive, yielding markedly improved performance: R2=0.973, RMSE = 0.211 μg cm−2, LOD = 0.716 μg cm−2, and LOQ = 2.169 μg cm−2. These results demonstrate that chemometric interval selection substantially enhances sensitivity and chemical interpretability, establishing handheld FTIR-interval partial least squares as an efficient, nondestructive alternative for routine pharmaceutical cleaning verification.


Downloadable publication

This is an electronic reprint of the original article.
This reprint may differ from the original in pagination and typographic detail. Please cite the original version.




Funding information in the publication
This research was conducted as part of the LifeFactFuture (LFF) consortium, a collaboration between the University of Turku, Orion Pharma, and leading Finnish companies in the life sciences and technology sectors. This work was supported by Business Finland under Grant 6819/31/2023.


Last updated on 14/08/2026 11:55:54 AM