A1 Vertaisarvioitu alkuperäisartikkeli tieteellisessä lehdessä

New Artisse intrasaccular device for intracranial aneurysm treatment: short term clinical and angiographic result from the prospective registry INSPIRE-A




TekijätHohenstatt, Sophia; Costalat, Vincent; Dargazanli, Cyril; Killer-Oberpfalzer, Monika; Schreiber, Barbara; Rautio, Riitta; Sinisalo, Matias; Lamin, Saleh; Chew, Han Seng; Spelle, Laurent; Tomasello, Alejandro; Patankar, Tufail; Piano, Mariangela; Fiehler, Jens; Möhlenbruch, Markus A.

KustantajaBMJ PUBLISHING GROUP

KustannuspaikkaLONDON

Julkaisuvuosi2025

JournalJournal of NeuroInterventional Surgery

Tietokannassa oleva lehden nimiJOURNAL OF NEUROINTERVENTIONAL SURGERY

Lehden akronyymiJ NEUROINTERV SURG

Sivujen määrä10

ISSN1759-8478

eISSN1759-8486

DOIhttps://doi.org/10.1136/jnis-2024-022576

Verkko-osoitehttps://doi.org/10.1136/jnis-2024-022576

Rinnakkaistallenteen osoitehttps://research.utu.fi/converis/portal/detail/Publication/485206166


Tiivistelmä

Background. Intrasaccular devices have broadened treatment options for wide necked aneurysms. This study presents the preliminary experience with the Artisse 2.0 device.

Methods. Innovative NeurovaScular Product SurveIllance REgistry (INSPIRE) is a non-randomized, multicenter, real world clinical study with treatment arms for aneurysms (INSPIRE-A) and acute ischemic stroke (INSPIRE-S). This interim analysis included 87 patients enrolled from November 2022 to April 2024 in the INSPIRE-A Artisse cohort across 16 Europoean centers. Procedures followed standard clinical care, with 6 months of follow-up. Safety and efficacy endpoints included major stroke, neurological death, serious adverse events (SAEs), aneurysm occlusion, and retreatment rates. An independent core laboratory assessed imaging, and all SAEs were reviewed by a clinical events committee. The Artisse steering committee provided independent oversight of the data.

Results. The Artisse device achieved an overall successful implantation rate of 96.6% (84/87), with satisfactory placement rates of 98.7% (74/75) for unruptured and 88.9% (8/9) for ruptured aneurysms. Following the procedure, 46.2% of unruptured aneurysm patients were receiving antiplatelet therapy (APT), predominantly aspirin monotherapy, while no ruptured aneurysm patients received APT. Device related SAE rate was 1.3% (1/87), and the overall stroke rate was 2.3% (2/87), including both ruptured and unruptured aneurysms. At 6 months, 80.0% (28/35) of patients with unruptured aneurysms showed complete obliteration, with no recurrences or retreatments.

Conclusions. Preliminary experience with the Artisse 2.0 device demonstrated high technical success, favorable safety, and efficacy in aneurysm obliteration at 6 months. Larger studies with longer follow-up periods are needed to confirm these findings.


Ladattava julkaisu

This is an electronic reprint of the original article.
This reprint may differ from the original in pagination and typographic detail. Please cite the original version.




Julkaisussa olevat rahoitustiedot
Innovative NeurovaScular Product SurveIllance REgistry (INSPIRE)-A isvsponsored by Medtronic


Last updated on 2025-13-05 at 17:55