A1 Vertaisarvioitu alkuperäisartikkeli tieteellisessä lehdessä

Targeted and Untargeted Amine Metabolite Quantitation in Single Cells with Isobaric Multiplexing




TekijätHeininen, Juho; Movahedi, Parisa; Kotiaho, Tapio; Kostiainen, Risto; Pahikkala, Tapio; Teppo, Jaakko

KustantajaWiley-VCH

Julkaisuvuosi2024

JournalChemistry - A European Journal

Lehden akronyymiChemistry

Artikkelin numeroe202403278

Vuosikerta30

Numero72

ISSN0947-6539

eISSN1521-3765

DOIhttps://doi.org/10.1002/chem.202403278

Verkko-osoitehttps://doi.org/10.1002/chem.202403278

Rinnakkaistallenteen osoitehttps://research.utu.fi/converis/portal/detail/Publication/459037549


Tiivistelmä

We developed a single cell amine analysis approach utilizing isobarically multiplexed samples of 6 individual cells along with analyte abundant carrier. This methodology was applied for absolute quantitation of amino acids and untargeted relative quantitation of amines in a total of 108 individual cells using nanoflow LC with high-resolution mass spectrometry. Together with individually determined cell sizes, this provides quantification of intracellular concentrations within individual cells. The targeted method was partially validated for 10 amino acids with limits of detection in low attomoles, linear calibration range covering analyte amounts typically from 30 amol to 120 fmol, and correlation coefficients (R) above 0.99. This was applied with cell sizes recorded during dispensing to determine millimolar intracellular amino acid concentrations. The untargeted approach yielded 249 features that were detected in at least 25% of the single cells, providing modest cell type separation on principal component analysis. Using Greedy forward selection with regularized least squares, a sub-selection of 100 features explaining most of the difference was determined. These features were annotated using MS2 from analyte standards and accurate mass with library search. The approach provides accessible, sensitive, and high-throughput method with the potential to be expanded also to other forms of ultrasensitive analysis.


Ladattava julkaisu

This is an electronic reprint of the original article.
This reprint may differ from the original in pagination and typographic detail. Please cite the original version.




Julkaisussa olevat rahoitustiedot
Research Council of Finland (project #321472) is acknowledged for funding.