The out-of-plane contact shield and mA-modulation – the effect on fetal absorbed dose




TekijätLarjava, Heli R.S.; Eneh, Chibuzor T.M.; Saikkonen, Aleksi; Parkkola, Riitta K.

KustantajaOxford University Press (OUP)

KustannuspaikkaOXFORD

Julkaisuvuosi2024

JournalRadiation Protection Dosimetry

Tietokannassa oleva lehden nimiRadiation Protection Dosimetry

Lehden akronyymiRADIAT PROT DOSIM

Vuosikerta200

Numero19

Aloitussivu1945

Lopetussivu1949

Sivujen määrä5

ISSN0144-8420

eISSN1742-3406

DOIhttps://doi.org/10.1093/rpd/ncae204

Verkko-osoitehttps://doi.org/10.1093/rpd/ncae204

Rinnakkaistallenteen osoitehttps://research.utu.fi/converis/portal/detail/Publication/458946236


Tiivistelmä
The effect of patient shielding on fetal radiation dose was evaluated in computed tomography pulmonary angiography with the out-of-plane shield visible in the localizer but absent in the scan range in chest computed tomography (CT). An anthropomorphic phantom with additional prosthetic pregnancy belly was scanned with different CT scanners using clinical imaging protocols and radiophotoluminescence dosemeters (GD-352 M). The out-of-plane shield decreased the fetal absorbed radiation dose with Siemens Somatom go.Up, Canon Aquilion Prime SP and Canon Aquilion One scanners. The decrease was 3.9%-39.4% (0.01-0.09 mGy). With GE Optima the shield increased the fetal dose by 100% (0.23 mGy), with Canon Aquilion One and GE Optima scanners the abdomen dose increase was 17.5% and 36.4%, respectively (0.61 and 1.38 mGy). Applying an out-of-plane shield outside the scanned volume may increase the fetal radiation dose during CT when using tube current modulation, depending on the make and model of the CT scanner.

Ladattava julkaisu

This is an electronic reprint of the original article.
This reprint may differ from the original in pagination and typographic detail. Please cite the original version.




Julkaisussa olevat rahoitustiedot
This work was supported by the State Research Funding of the Turku University Hospital.


Last updated on 2025-27-01 at 20:04