Polymerization shrinkage stress of contemporary dental composites: Comparison of two measurement methods




TekijätSzczesio-Wlodarczyk Agata, Garoushi Sufyan, Vallittu Pekka, Bociong Kinga, Lassila Lippo

KustantajaJapanese Society for Dental Materials and Devices

Julkaisuvuosi2024

JournalDental Materials Journal

Vuosikerta43

Numero2

Aloitussivu155

Lopetussivu163

ISSN0287-4547

eISSN1881-1361

DOIhttps://doi.org/10.4012/dmj.2023-192

Verkko-osoitehttps://doi.org/10.4012/dmj.2023-192

Rinnakkaistallenteen osoitehttps://research.utu.fi/converis/portal/detail/Publication/381030511


Tiivistelmä

The aim of this study was to compare two testing methodologies employed for assessing the polymerization shrinkage stress of dental resin composites. Ten commercial resin composites were investigated (EverX Posterior & Flow; G-ænial Anterior, Posterior, A'CHORD & Universal Injectable; Filtek One Bulk Fill & Universal Restorative; SDR flow+ and Aura Bulk Fill). Photoelastic and contraction forces measurement methods were performed. The slope of the linear trendline and C-factor of specimens were calculated. The shrinkage stress values (range between: 6.4-13.4 MPa) obtained by the photoelastic method were higher for all resin composites than the values obtained by contraction forces measurements (range between 1.2-4.8 MPa). However, there was a strong linear correlation between these methods (r=0.8). The use of both investigated methods revealed important information about the shrinkage behavior of the restorative resin composites.


Ladattava julkaisu

This is an electronic reprint of the original article.
This reprint may differ from the original in pagination and typographic detail. Please cite the original version.




Ladattava julkaisu

This is an electronic reprint of the original article.
This reprint may differ from the original in pagination and typographic detail. Please cite the original version.





Last updated on 2025-21-03 at 10:12