Validation of indirect calorimetry for measurement of energy expenditure in healthy volunteers undergoing pressure controlled non-invasive ventilation support




TekijätSiirala W, Noponen T, Olkkola KT, Vuori A, Koivisto M, Hurme S, Aantaa R

KustantajaSPRINGER HEIDELBERG

Julkaisuvuosi2012

JournalJournal of Clinical Monitoring and Computing

Tietokannassa oleva lehden nimiJOURNAL OF CLINICAL MONITORING AND COMPUTING

Lehden akronyymiJ CLIN MONIT COMPUT

Numero sarjassa1

Vuosikerta26

Numero1

Aloitussivu37

Lopetussivu43

Sivujen määrä7

ISSN1387-1307

DOIhttps://doi.org/10.1007/s10877-011-9331-z

Rinnakkaistallenteen osoitehttps://research.utu.fi/converis/portal/detail/Publication/3555154


Tiivistelmä
The aim of this validation study was to assess the reliability of gas exchange measurement with indirect calorimetry among subjects who undergo non-invasive ventilation (NIV). Oxygen consumption (VO2) and carbon dioxide production (VCO2) were measured in twelve healthy volunteers. Respiratory quotient (RQ) and resting energy expenditure (REE) were then calculated from the measured VO2 and VCO2 values. During the measurement period the subjects were breathing spontaneously and ventilated using NIV. Two different sampling air flow values 40 and 80 l/min were used. The gas leakage from the measurement setup was assessed with a separate capnograph. The mean weight of the subjects was 93 kg. Their mean body mass index was 29 (range 22-40) kg/m(2). There was no statistically significant difference in the measured values for VO2, VCO2, RQ and REE during NIV-supported breathing and spontaneous breathing. The change of sampling air flow had no statistically significant effect on any of the above parameters. We found that REE can be accurately measured with an indirect calorimeter also during NIV-supported breathing and the change of sampling air flow does not distort the gas exchange measurement. A higher sampling air flow in indirect calorimetry decreases the possibility for air leakages in the measurement system and increases the reliability of REE measurement.

Ladattava julkaisu

This is an electronic reprint of the original article.
This reprint may differ from the original in pagination and typographic detail. Please cite the original version.





Last updated on 2024-26-11 at 12:49