A1 Vertaisarvioitu alkuperäisartikkeli tieteellisessä lehdessä
Load-bearing capacity of fiber-reinforced composite abutments and one-piece implants
Tekijät: M Etxeberria, A A Abdulmajeed, T Escuin, M Vinas, L V Lassila, T O Närhi
Julkaisuvuosi: 2015
Journal: European Journal of Prosthodontics and Restorative Dentistry
Vuosikerta: 23
Numero: 2
Aloitussivu: 62
Lopetussivu: 69
Sivujen määrä: 8
ISSN: 0965-7452
DOI: https://doi.org/10.1992/EJPRD_1434Etxeberria08
Verkko-osoite: http://www.ejprd.org/download.php?op=institution_view_article&article_id=760
Fiber-reinforced composites (FRC) can potentially help in a physiologic stress transmission
due to its excellent biomechanical matching with living tissues. Novel one-piece
FRC implants and abutments with two different fiber orientations were loaded until failure
to assess the load-bearing capacity, fracture patterns, and precision of fit. The onepiece
FRC implants showed significantly higher load-bearing capacity compared to FRC
abutments regardless of the fiber orientation (p<0.001). For FRC abutments, bidirectional
abutments showed significantly higher loads compared to unidirectional abutments
(p<0.001). The type of structure and fiber orientation are strong determinant factors of
the load-bearing capacity of FRC implants and abutments.