A1 Vertaisarvioitu alkuperäisartikkeli tieteellisessä lehdessä

Optimization of multi-photon event discrimination levels using Poisson statistics




TekijätSoukka J., Virkki A., Hänninen P., Soini J.

Julkaisuvuosi2004

Lehti:Optics Express

Tietokannassa oleva lehden nimiOptics Express

Vuosikerta12

Numero1

Aloitussivu84

Lopetussivu89

Sivujen määrä6

ISSN1094-4087

DOIhttps://doi.org/10.1364/OPEX.12.000084

Verkko-osoitehttp://api.elsevier.com/content/abstract/scopus_id:2942717048


Tiivistelmä
In applications where random multi-photon events must be distinguishable from the background, detection of the signals must be based on either analog current measurement or photon counting and multi-level discrimination of single and multi-photon events. In this paper a novel method for optimizing photomultiplier (PMT) pulse discrimination levels in single- and multi-photon counting is demonstrated. This calibration method is based on detection of photon events in coincidence to short laser pulses. The procedure takes advantage of Poisson statistics of single- and multi-photon signals and it is applicable to automatic calibration of photon counting devices on production line. Results obtained with a channel photomultiplier (CPM) are shown. By use of three parallel discriminators and setting the discriminator levels according to the described method resulted in a linear response over wide range of random single- and multi-photon signals. © 2004 Optical Society of America.