A1 Vertaisarvioitu alkuperäisartikkeli tieteellisessä lehdessä
Pitfalls of femoral Titanium Elastic Nailing
Tekijät: Salonen A, Lahdes-Vasama T, Mattila VM, Valipakka J, Pajulo O
Kustantaja: SAGE PUBLICATIONS LTD
Julkaisuvuosi: 2015
Journal: Scandinavian Journal of Surgery
Tietokannassa oleva lehden nimi: SCANDINAVIAN JOURNAL OF SURGERY
Lehden akronyymi: SCAND J SURG
Vuosikerta: 104
Numero: 2
Aloitussivu: 121
Lopetussivu: 126
Sivujen määrä: 6
ISSN: 1457-4969
DOI: https://doi.org/10.1177/1457496914529275
Conclusions: Based on this study, two types of pitfalls in a small volume center were found. Titanium elastic nail ends were left unbent and too long. We recommend palpating the nail ends to exclude nail prominence and to verify free movement of the knee after nail cutting and bending. Fracture instability was caused by inserting titanium elastic nailing-nails that were too narrow. To avoid this complication, careful preoperative planning to select the proper-size titanium elastic nailing-nails and intraoperative testing of fracture stability under continuous fluoroscopy after the operation is advised.