Design and Implementation of a Non-invasive and Cuff-less Arterial Blood Pressure Monitoring System




Seyed Mohsen Anvari, Mohammadreza Yazdchi, Amirhossein Kayvanpour, Seyed Mohammad Hasan Nayebpour, Tero Koivisto, Mojtaba Jafari Tadi

Christine Pickett, Cristiana Corsi, Pablo Laguna, Rob MacLeod

Computing in Cardiology

2017

Computing in Cardiology

Computing in Cardiology

44

4

2325-8861

DOIhttps://doi.org/10.22489/CinC.2017.325-471

http://www.cinc.org/archives/2017/pdf/325-471.pdf

https://research.utu.fi/converis/portal/detail/Publication/28771555



Hypertension or high blood pressure (BP) is one of the most common worldwide disease leading to heart attack or stroke. Continuous assessment of blood pressure level is key to diagnosing hypertension. In this study, we de- signed and tested a dedicated cuff-less monitoring system which estimates BP level without need for calibration. We obtained continuous measurements from 40 healthy sub- jects (30 males and 10 females) ranging from 20-30 years old. Our measurement protocol consisted of 15 minutes si- multaneous electrocardiography (ECG) and photoplethys- mography (PPG) within three sessions, i.e. rest, bicycle exercise, and recovery. From ECG and PPG signals, we obtained 34 candidate features from which up to 9 features were selected to estimate systolic and diastolic BP levels. We validate our results with three regression models such as linear regression, support vector machines (SVM) re- gression, and multilayer perceptron (MLP) to obtain the best results. The study provides a promising approach for modern cuff-less BP monitoring devices. 