A1 Refereed original research article in a scientific journal

Analysis of F-18-labelled synthesis products on TLC plates: Comparison of radioactivity scanning, film auto radiography, and a phosphoimaging technique




AuthorsKamarainen EL, Haaparanta M, Siitari-Kauppi M, Koivula T, Lipponen T, Solin O

PublisherPERGAMON-ELSEVIER SCIENCE LTD

Publication year2006

JournalApplied Radiation and Isotopes

Journal name in sourceAPPLIED RADIATION AND ISOTOPES

Journal acronymAPPL RADIAT ISOTOPES

Volume64

Issue9

First page 1043

Last page1047

Number of pages5

ISSN0969-8043

DOIhttps://doi.org/10.1016/j.apradiso.2006.04.012


Abstract
We compared radioactivity scanning, film autoradiography, and digital photostimulated luminescence (PSL) autoradiography (phosphoimaging technique) in detection of radioactivity on thin-layer chromatography (TLC) plates. TLC combined with radioactivity detection is rapid, simple, and relatively flexible. Here, F-18-labelled synthesis products were analyzed by TLC and the radioactivity distribution on the plates determined using the three techniques. Radioactivity scanning is appropriate only with good chromatographic resolution and previously validated scanning parameters. Film autoradiography exhibits poor linearity if radioactivity varies greatly. PSL provides high sensitivity and resolution and superior linearity compared with the other methods. (c) 2006 Elsevier Ltd. All rights reserved.