A1 Refereed original research article in a scientific journal

The FRESH Study: Treatment of Intracranial Aneurysms with the New FRED X Flow Diverter with Antithrombotic Surface Treatment Technology?First Multicenter Experience in 161 Patients




AuthorsVollherbst D. F., Luecking H., DuPlessis J., Sonnberger M., Maurer C., Kocer N., Killer-Oberpfalzer M., Rautio R., Valvassori L., Berlis A., Gasser S., Gatt S., Doerfler A., Bendszus M., Moehlenbruch M. A.

PublisherAMER SOC NEURORADIOLOGY

Publication year2023

Journal: American Journal of Neuroradiology

Journal name in sourceAMERICAN JOURNAL OF NEURORADIOLOGY

Journal acronymAM J NEURORADIOL

Volume44

Issue4

First page 474

Last page480

Number of pages7

ISSN0195-6108

eISSN1936-959X

DOIhttps://doi.org/10.3174/ajnr.A7834

Publication's open availability at the time of reportingNo Open Access

Publication channel's open availability Partially Open Access publication channel

Web address https://www.ajnr.org/content/44/4/474


Abstract

BACKGROUND AND PURPOSE: Flow diverters with antithrombotic coatings are increasingly used to improve the safety of flow diverter treatments of intracranial aneurysms. This study aimed to investigate the safety and short-term efficacy of the new FRED X flow diverter.

MATERIALS AND METHODS: Medical charts and procedural and imaging data of a consecutive series of patients with intracranial aneurysms who were treated with the FRED X at 9 international neurovascular centers were retrospectively analyzed.

RESULTS: One hundred sixty-one patients (77.6% women; mean age, 55?years) with 184 aneurysms (11.2% acutely ruptured) were included in this study. Most aneurysms were located in the anterior circulation (77.0%), most frequently at the ICA (72.7%). The FRED X was successfully implanted in all procedures. Additional coiling was performed in 29.8%. In-stent balloon angioplasty was necessary in 2.5%. The rate of major adverse events was 3.1%. Thrombotic events occurred in 7 patients (4.3%) with 4 intra- and 4 postprocedural in-stent thromboses, respectively (1 patient had both peri- and postprocedural thrombosis). Of these thrombotic events, only 2 (1.2%) led to major adverse events (ischemic strokes). Postinterventional neurologic morbidity and mortality were observed in 1.9% and 1.2%, respectively. The rate of complete aneurysm occlusion after a mean follow-up of 7.0?months was 66.0%.

CONCLUSIONS: The new FRED X is a safe and feasible device for aneurysm treatment. In this retrospective multicenter study, the rate of thrombotic complications was low, and the short-term occlusion rates are satisfactory.