A1 Vertaisarvioitu alkuperäisartikkeli tieteellisessä lehdessä
Inter-Laboratory Study on Measuring the Surface Charge of Electrically Polarized Hydroxyapatite
Tekijät: Ubele-Kalnina Darta, Nakamura Miho, Gross Karlis Agris
Kustantaja: MDPI
Julkaisuvuosi: 2023
Journal: Journal of Functional Biomaterials
Tietokannassa oleva lehden nimi: Journal of Functional Biomaterials
Artikkelin numero: 100
Vuosikerta: 14
Numero: 2
ISSN: 2079-4983
eISSN: 2079-4983
DOI: https://doi.org/10.3390/jfb14020100
Verkko-osoite: https://doi.org/10.3390/jfb14020100
Rinnakkaistallenteen osoite: https://research.utu.fi/converis/portal/detail/Publication/179123336
Surface charges on implants improve integration into bone and so require a clear protocol for achieving a surface charge and comparable results from different laboratories. This study sintered hydroxyapatite (HAp) at one laboratory to remove the influence of the microstructure on surface charge and then polarized/depolarized the pellets at two different laboratories (in Tokyo and Riga). Surface charges on HAp pellets induced by electric polarization at 400 °C in a 5 kV/cm DC electric field were measured by the thermally stimulated depolarization current (TSDC) method as 6–9 µC/cm2. The surface charge results were comparable between laboratories and also agreed with previously documented values. Recommendations describe conditions for polarization and depolarization to generate a surface charge and repeatedly achieve a comparable outcome. A visual display of the polarization mechanisms and the contribution to surface charge point to further aspects that need further development.
Ladattava julkaisu This is an electronic reprint of the original article. |