Evaluation of three methods for delineation and attenuation estimation of the sinus region in MR-based attenuation correction for brain PET-MR imaging




TekijätLindén Jani, Teuho Jarmo, Teräs Mika, Klén Riku

KustantajaBMC

Julkaisuvuosi2022

JournalBMC Medical Imaging

Tietokannassa oleva lehden nimiBMC MEDICAL IMAGING

Lehden akronyymiBMC MED IMAGING

Artikkelin numero 48

Vuosikerta22

Sivujen määrä14

ISSN1471-2342

eISSN1471-2342

DOIhttps://doi.org/10.1186/s12880-022-00770-0

Verkko-osoitehttps://bmcmedimaging.biomedcentral.com/articles/10.1186/s12880-022-00770-0

Rinnakkaistallenteen osoitehttps://research.utu.fi/converis/portal/detail/Publication/175003069


Tiivistelmä

Background

Attenuation correction is crucial in quantitative positron emission tomography-magnetic resonance (PET-MRI) imaging. We evaluated three methods to improve the segmentation and modelling of the attenuation coefficients in the nasal sinus region. Two methods (cuboid and template method) included a MRI-CT conversion model for assigning the attenuation coefficients in the nasal sinus region, whereas one used fixed attenuation coefficient assignment (bulk method).

Methods

The study population consisted of data of 10 subjects which had undergone PET-CT and PET-MRI. PET images were reconstructed with and without time-of-flight (TOF) using CT-based attenuation correction (CTAC) as reference. Comparison was done visually, using DICE coefficients, correlation, analyzing attenuation coefficients, and quantitative analysis of PET and bias atlas images.

Results

The median DICE coefficients were 0.824, 0.853, 0.849 for the bulk, cuboid and template method, respectively. The median attenuation coefficients were 0.0841 cm-1, 0.0876 cm-1, 0.0861 cm-1 and 0.0852 cm-1, for CTAC, bulk, cuboid and template method, respectively. The cuboid and template methods showed error of less than 2.5% in attenuation coefficients. An increased correlation to CTAC was shown with the cuboid and template methods. In the regional analysis, improvement in at least 49% and 80% of VOI was seen with non-TOF and TOF imaging. All methods showed errors less than 2.5% in non-TOF and less than 2% in TOF reconstructions.

Conclusions

We evaluated two proof-of-concept methods for improving quantitative accuracy in PET/MRI imaging and showed that bias can be further reduced by inclusion of TOF. Largest improvements were seen in the regions of olfactory bulb, Heschl's gyri, lingual gyrus and cerebellar vermis. However, the overall effect of inclusion of the sinus region as separate class in MRAC to PET quantification in the brain was considered modest.