A1 Refereed original research article in a scientific journal

Myocardial perfusion quantification with Rb-82 PET: good interobserver agreement of Carimas software on global, regional, and segmental levels




AuthorsNesterov Sergey V, Deshayes Emmanuel, Juarez-Orozco Luis Eduardo, deKemp Robert A, Sciagrà Roberto, Malaspina Simona, Settimo Leonardo, Han Chunlei, Ryzhkova Darja V, Kostina Irina S, Gwet Kilem L, Prior John O, Knuuti Juhani M

PublisherSpringer

Publication year2022

JournalAnnals of Nuclear Medicine

Journal name in sourceANNALS OF NUCLEAR MEDICINE

Journal acronymANN NUCL MED

Number of pages8

ISSN0914-7187

eISSN1864-6433

DOIhttps://doi.org/10.1007/s12149-022-01729-5

Web address https://link.springer.com/article/10.1007/s12149-022-01729-5

Self-archived copy’s web addresshttps://research.utu.fi/converis/portal/detail/Publication/174852692


Abstract

Purpose
To estimate the interobserver agreement of the Carimas software package (SP) on global, regional, and segmental levels for the most widely used myocardial perfusion PET tracer-Rb-82.

Materials and methods
Rest and stress Rb-82 PET scans of 48 patients with suspected or known coronary artery disease (CAD) were analyzed in four centers using the Carimas SP. We considered values to agree if they simultaneously had an intraclass correlation coefficient (ICC) > 0.75 and a difference < 20% of the median across all observers.

Results
The median values on the segmental level were 1.08 mL/min/g for rest myocardial blood flow (MBF), 2.24 mL/min/g for stress MBF, and 2.17 for myocardial flow reserve (MFR). For the rest MBF and MFR, all the values at all the levels fulfilled were in excellent agreement. For stress MBF, at the global and regional levels, all the 24 comparisons showed excellent agreement. Only 1 out of 102 segmental comparisons (seg. 14) was over the adequate agreement limit-23.5% of the median value (ICC = 0.95).

Conclusion
Interobserver agreement for Rb-82 PET myocardial perfusion quantification analyzed with Carimas is good at any LV segmentation level-global, regional, and segmental. It is good for all the estimates-rest MBF, stress MBF, and MFR.


Downloadable publication

This is an electronic reprint of the original article.
This reprint may differ from the original in pagination and typographic detail. Please cite the original version.





Last updated on 2024-26-11 at 20:35