A1 Refereed original research article in a scientific journal

Reporting Sysmex XN Absolute Neutrophil Count in Samples with Leukocyte Analyzer Flagging




AuthorsLinko-Parvinen AM, Turkia H

PublisherOxford University Press

Publication year2021

JournalLaboratory Medicine -Philadelphia then Chicago-

Journal name in sourceLaboratory medicine

Journal acronymLab Med

ISSN0007-5027

eISSN1943-7730

DOIhttps://doi.org/10.1093/labmed/lmaa058


Abstract
To provide faster laboratory data reporting, we evaluated the accuracy of Sysmex XN (Sysmex Inc, Kobe, Japan) absolute neutrophil count (ANC) in the presence of analyzer flagging.Sysmex XN and manual microscopy ANC were compared with 80 autovalidated control specimens and with 280 study specimens with analyzer flagging regarding immature granulocytes (IG) >3% or other leukocyte abnormalities. Specimens with ambiguous neutrophil clusters were excluded.A slight positive overall method bias was seen for Sysmex XN compared to manual microscopy (n = 280), 0.025 (95% confidence interval [CI], -0.023 to 0.069) × 109/L. With IG > 10% (n = 123) the bias was larger, but not clinically significant, 0.17 (95% CI, 0.060-0.25) × 109/L. No clinically significant difference was seen in neutropenic (ANC < 1.5 × 109/L) specimens (n = 91), 0.070 (95% CI, -0.013 to 0.14) × 109/L.These data indicate that Sysmex XN ANC can be reported in the presence of certain analyzer flagging to improve patient care.



Last updated on 2024-26-11 at 15:44