A1 Vertaisarvioitu alkuperäisartikkeli tieteellisessä lehdessä

Feasibility of Multiparameter MRI-Guided Percutaneous Biopsy for Central Lung Lesions With Atelectasis




TekijätLi Peipei; Li Chengli; Xu Yujun; He Xiangmeng; Sequeiros Roberto Blanco; Liu Ming

KustantajaThe Korean Society of Radiology

KustannuspaikkaSEOUL

Julkaisuvuosi2025

JournalKorean Journal of Radiology

Tietokannassa oleva lehden nimiKorean Journal of Radiology

Lehden akronyymiKOREAN J RADIOL

Vuosikerta26

Numero5

Aloitussivu498

Lopetussivu507

Sivujen määrä10

ISSN1229-6929

eISSN2005-8330

DOIhttps://doi.org/10.3348/kjr.2024.0818

Verkko-osoitehttps://doi.org/10.3348/kjr.2024.0818

Rinnakkaistallenteen osoitehttps://research.utu.fi/converis/portal/detail/Publication/492275262


Tiivistelmä
Objective: To prospectively evaluate the feasibility, accuracy, and safety of multiparameter MRI-guided percutaneous biopsy using a 1T open MRI scanner for evaluating suspicious centrally located lung lesions with associated post-obstructive atelectasis. Materials and Methods: In this single-center study, MRI-guided percutaneous coaxial cutting biopsy was performed for 107 suspicious central lung lesions with associated post-obstructive atelectasis in 107 patients between July 2015 and December 2020. A fast T2-weighted imaging (T2WI)-turbo spin echo (TSE) sequence and an enhanced fast T1-weighted imaging (T1WI)-TSE sequence were used to identify, localize, and biopsy lung lesions, and diffusion-weighted imaging (DWI) was used as a supplementary sequence for identifying the lesion location. The final diagnosis was confirmed by surgical histopathology or clinical follow-up for a minimum of 24 months. The sensitivity, specificity, and accuracy for diagnosing lung malignancies were calculated, and the complications were recorded for each case. Results: Using multiparameter MRI, central lung lesions could be clearly distinguished from post-obstructive atelectasis in 96 patients (89.7%). The sensitivity, specificity, and accuracy of MRI-guided percutaneous biopsy for diagnosing lung malignancy was 97.0% (98/101), 100% (6/6), and 97.2% (104/107), respectively. Self-limited hemoptysis occurred in three patients. Pneumothorax occurred in five patients, of which none required pleural drainage. No serious procedure-related complications were observed. Conclusion: As a technology that does not involve ionizing radiation, multiparameter MRI-guided percutaneous coaxial cutting biopsy is a safe and accurate diagnostic technique for evaluating centrally located lung lesions associated with post-obstructive atelectasis.

Ladattava julkaisu

This is an electronic reprint of the original article.
This reprint may differ from the original in pagination and typographic detail. Please cite the original version.




Julkaisussa olevat rahoitustiedot
This research has been supported by Natural Science Foundation of Shandong Province (ZR2020MH228).