Vertaisarvioitu artikkeli konferenssijulkaisussa (A4)

An Automated Device for Recording Peripheral Arterial Waveform




Julkaisun tekijät: Tuukka Panula, Juuso Blomster, Mikko Pänkäälä, Tero Koivisto, Matti Kaisti

Konferenssin vakiintunut nimi: Computing in Cardiology

Julkaisuvuosi: 2019

Journal: Computing in Cardiology

Kirjan nimi *: CinC 2019

Sarjan nimi: Computing in Cardiology

ISSN: 2325-8861

DOI: http://dx.doi.org/10.22489/cinc.2019.260

Rinnakkaistallenteen osoite: https://research.utu.fi/converis/portal/detail/Publication/45017924


Tiivistelmä

The aim of the study was to develop an automated device for recording peripheral arterial pulse wave, in order to assess cardiovascular health. Recent studies have shown that photoplethysmography (PPG) is a viable technique to measure peripheral pressure waveform. We developed a small motorized device that can measure pulse waveform from a finger. The device targets the distal transverse palmar arch (DTPA) artery using infrared wavelength PPG. Measurements were taken from healthy subjects (n = 8).The device was validated by performing HR detection and waveform analysis. The device was able to record high quality blood pressure calibrated arterial waveforms and detect beat-to-beat heart rate allowing the assessment of cardiovascular health status.


Ladattava julkaisu

This is an electronic reprint of the original article.
This reprint may differ from the original in pagination and typographic detail. Please cite the original version.




Last updated on 2022-07-04 at 17:38