Clinical evaluation of in silico planning and real-time simulation of hepatic radiofrequency ablation (ClinicIMPPACT Trial)
Authors: Michael Moche, Harald Busse, Jurgen J. Futterer, Camila A. Hinestrosa, Daniel Seider, Philipp Brandmaier, Marina Kolesnik, Sjoerd Jenniskens, Roberto Blanco Sequeiros, Gaber Komar, Mika Pollari, Martin Eibisberger, Horst Rupert Portugaller, Philip Voglreiter, Ronan Flanagan, Panchatcharam Mariappan, Martin Reinhardt
Publisher: Springer Verlag
Publication year: 2020
Journal: European Radiology
Journal name in source: European Radiology
Volume: 30
Issue: 2
First page : 934
Last page: 942
eISSN: 1432-1084
DOI: https://doi.org/10.1007/s00330-019-06411-5
Objectives
To evaluate the accuracy and clinical integrability of a comprehensive simulation tool to plan and predict radiofrequency ablation (RFA) zones in liver tumors.
MethodsForty-five patients with 51 malignant hepatic lesions of different origins were included in a prospective multicenter trial. Prior to CT-guided RFA, all patients underwent multiphase CT which included acquisitions for the assessment of liver perfusion. These data were used to generate a 3D model of the liver. The intra-procedural position of the RFA probe was determined by CT and semi-automatically registered to the 3D model. Size and shape of the simulated ablation zones were compared with those of the thermal ablation zones segmented in contrast-enhanced CT images 1 month after RFA; procedure time was compared with a historical control group.
ResultsSimulated and segmented ablation zone volumes showed a significant correlation (ρ = 0.59, p < 0.0001) and no significant bias (Wilcoxon’s Z = 0.68, p = 0.25). Representative measures of ablation zone comparison were as follows: average surface deviation (absolute average error, AAE) with 3.4 ± 1.7 mm, Dice similarity coefficient 0.62 ± 0.14, sensitivity 0.70 ± 0.21, and positive predictive value 0.66 ± 0. There was a moderate positive correlation between AAE and duration of the ablation (∆t; r = 0.37, p = 0.008). After adjustments for inter-individual differences in ∆t, liver perfusion, and prior transarterial chemoembolization procedures, ∆t was an independent predictor of AAE (ß = 0.03 mm/min, p = 0.01). Compared with a historical control group, the simulation added 3.5 ± 1.9 min to the procedure.
ConclusionThe validated simulation tool showed acceptable speed and accuracy in predicting the size and shape of hepatic RFA ablation zones. Further randomized controlled trials are needed to evaluate to what extent this tool might improve patient outcomes.