The Choice of Home Blood Pressure Monitoring Device Matters – Differences between Three Commercial Monitors Used in Finnish General Practice




Differences between Three Commercial Monitors Used in Finnish General Practice

Varis JP, Kantola I

2013

Journal of General Practice

106

1

1

1

2329-9126

DOIhttps://doi.org/10.4172/2329-9126.1000106



Objectives: The use of home blood pressure measurement in the follow-up of antihypertensive treatment is rapidly increasing. Published non-commercial comparisons or validation studies between different types of home monitors have been rare. We describe here the results from a study in which three home blood pressure monitors from two manufacturers were compared.

Methods: Three different upper arm automatic blood pressure measuring devices: Microlife BP 3AC 1-1,

Microlife BP3AG1 and Omron M4-I were compared in a randomized study. It was carried out among 65 Finnish normo- and hypertensive people who used all three devices at home for one-week of blood pressure monitoring and measured their blood pressure twice both in the morning and in the evening.

Results: An about 2 mm Hg difference in systolic, diastolic and mean pressure was observed between Omron M4-I and the Microlife devices whereas the Microlife devices showed equal performance. A significant correlation in the mean pressure was observed between all three devices.

Conclusions: Although all these devices have passed the British Hypertension Society (BHS) qualification, the blood pressure monitoring results in this study showed that there may be a small blood pressure difference between the devices from separate manufacturers. Blood pressure monitoring results are most reliable when the same device is used consistently.