CHOLESTEROL MEASUREMENT BY REFLOTRON DRY CHEMISTRY IN INFANTS, CHILDREN AND ADULTS




TekijätLAPINLEIMU H, VIIKARI J, JOKINEN E, ROUTI T, SALO P, HOLLO O, NIINIKOSKI H, RONNEMAA T, LEINO A, VALIMAKI I, SIMELL O

KustantajaSCANDINAVIAN UNIVERSITY PRESS

Julkaisuvuosi1994

JournalScandinavian Journal of Clinical and Laboratory Investigation

Tietokannassa oleva lehden nimiSCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION

Lehden akronyymiSCAND J CLIN LAB INV

Vuosikerta54

Numero1

Aloitussivu61

Lopetussivu65

Sivujen määrä5

ISSN0036-5513

DOIhttps://doi.org/10.3109/00365519409086511


Tiivistelmä

Cholesterol values measured with the Reflotron(R) dry chemistry (Boehringer Mannheim, Mannheim, Germany) and enzymatic (CHOD-PAP) routine method in 4150 venous blood samples from 580 infants aged 7 months, 1778. children aged 13 months to 19 years, and 1792 adults aged 18-63 years showed good correlation between the two assay systems (r > 0.92). However, the mean cholesterol concentrations were 4.5%, 3.7% and 2.3% lower in infants, children and adults, respectively, if measured with the Reflotron analyser as compared with values obtained with the CHOD-PAP method. These differences in the values were greater when cholesterol values were low (significance for the slope of regression line, p < 0.001). Values were below the detection limit of the Reflotron assay (2.59mmol l-1) in 104 (4.4%) and 4 (0.2%) of all paediatric and adult samples, respectively. When measured with the CHOD-PAP method, 75 (72%) and 3 (75%) of these same samples showed values which exceeded the 2.59mmol l-1 detection limit. We conclude that Reflotron dry chemistry analyser is precise in fast measurement of serum cholesterol both in all paediatric subjects and adults. A particular problem in paediatric age groups is that about 4% of children have values below the detection limit of the Reflotron analyser.




Last updated on 2024-26-11 at 16:31