A1 Vertaisarvioitu alkuperäisartikkeli tieteellisessä lehdessä
Three metal-on-metal hip replacements from the same manufacturer – A short- to midterm survival
Tekijät: Kostensalo I, Junnila M, Mokka J, Virolainen P, Vahlberg T, Mäkelä KT
Julkaisuvuosi: 2014
Journal: Acta Orthopaedica Belgica
Vuosikerta: 80
Numero: 2
Aloitussivu: 222
Lopetussivu: 227
Sivujen määrä: 6
ISSN: 0001-6462
The aim of this retrospective study was to evaluate short- to mid-term results of three different metal-on-metal hip devices from the same manufacturer. A total of 329 hip operations were performed in a single academic unit between 2004 and 2010 using either Birmingham hip resurfacing or Synergy--Birmingham and Synergy--R3 total hip arthroplasty. The overall survival rate at the end of the follow-up time for Birmingham hip resurfacing was 88%, for Synergy--Birmingham total hip arthroplasty 95%, and for Synergy--R3 total hip arthroplasty 81% (p = 0.036). Five revision operations were performed due to adverse reaction to metal debris. Head sizes > 50 mm had lower revision rates compared to smaller ones. Synergy--R3 had a poor survival already at short-term. The mid-term survival of Birmingham hip resurfacing arthroplasty was inferior compared to previous studies.