A dosimetric phantom study of dose accuracy and build-up effects using IMRT and RapidArc in stereotactic irradiation of lung tumours
Tekijät: Seppala J, Suilamo S, Kulmala J, Mali P, Minn H
Kustantaja: BIOMED CENTRAL LTD
Julkaisuvuosi: 2012
Journal: Radiation Oncology
Tietokannassa oleva lehden nimi: RADIATION ONCOLOGY
Lehden akronyymi: RADIAT ONCOL
Artikkelin numero: ARTN 79
Vuosikerta: 7
Aloitussivu: 79
Sivujen määrä: 10
ISSN: 1748-717X
DOI: https://doi.org/10.1186/1748-717X-7-79
Tiivistelmä
Conclusions: Overall, the peripheral doses of the simulated lung tumours were decreased by decreasing the SWA. To achieve adequate surface dose coverage to small lung tumours with a difference less than 1 mm in the isodose line radius between the open and modulated field, a larger than 6 mm SWA should be used in the dose delivery of SLRT.
Conclusions: Overall, the peripheral doses of the simulated lung tumours were decreased by decreasing the SWA. To achieve adequate surface dose coverage to small lung tumours with a difference less than 1 mm in the isodose line radius between the open and modulated field, a larger than 6 mm SWA should be used in the dose delivery of SLRT.