A1 Refereed original research article in a scientific journal

Measurements of [C-11]CO2 in Exhaled Air with a Positron-sensitive Single-wire Proportional Counter after i.v. Injection of [C-11]Acetate




AuthorsSipila HT, Heselius SL, Kalliokoski MK, Kemppainen J, Minn H, Tolvanen T, Roivainen A

Publication year2011

JournalIEEE conference record - Nuclear Science Symposium & Medical Imaging Conference

Journal name in source2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC)

Journal acronymIEEE NUCL SCI CONF R

First page 2397

Last page2399

Number of pages3

ISSN1082-3654


Abstract
A positron-sensitive, large-window single-wire proportional counter was used for measurements of [C-11]CO2 in exhaled air after i.v. injection of [C-11] acetate. The performance of the proportional counter was studied using Geant4 and Garfield simulation tools as well as experimentally by injections of known amounts of radioactive gas into the measuring volume. The proportional counter was found to be sensitive enough for measuring the low concentrations of [C-11]CO2 in the air exhaled by patients after i.v. injection of an C-11-Iabelled radiopharmaceutical. The limit of detection was 200 Bq/L for [C-11]CO2.



Last updated on 2024-26-11 at 17:47