Comparison of physical quality assurance between Scanora 3D and 3D Accuitomo 80 dental CT scanners




Ali AS, Fteita D, Kulmala J

PublisherCO-ACTION PUBLISHING

2015

Libyan Journal of Medicine

LIBYAN JOURNAL OF MEDICINE

LIBYAN J MED

ARTN 28038

10

12

1993-2820

1819-6357

DOIhttps://doi.org/10.3402/ljm.v10.28038



Conclusions: The radiation dose values in these two CBCT units are significantly less than those encountered in systemic CT scans. However, the dose seems to be affected more by changing the field of view rather than the voltage or amperage. The low doses were at the expense of the image quality produced, which was still acceptable. Although the spatial resolution and contrast were inferior to the medical images produced in systemic CT units, the present results recommend adopting CBCTs in maxillofacial imaging because of low radiation dose and adequate image quality.