A1 Vertaisarvioitu alkuperäisartikkeli tieteellisessä lehdessä
Survival of uncemented cups from a single manufacturer implanted from 1985 to 2013: Finnish Arthroplasty Register data
Tekijät: Antton Palomäki, Lasse Lempainen, Markus Matilainen, Antti Eskelinen, Ville Remes, Petri Virolainen, Keijo T. Mäkelä
Kustantaja: SPRINGER
Julkaisuvuosi: 2017
Journal: Archives of Orthopaedic and Trauma Surgery
Tietokannassa oleva lehden nimi: ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY
Lehden akronyymi: ARCH ORTHOP TRAUM SU
Vuosikerta: 137
Numero: 3
Aloitussivu: 311
Lopetussivu: 320
Sivujen määrä: 10
ISSN: 0936-8051
eISSN: 1434-3916
DOI: https://doi.org/10.1007/s00402-017-2624-z
Tiivistelmä
Total hip arthroplasty (THA) with uncemented implants has been a relatively common procedure in Finland during the recent decades. The most common brand of uncemented implants was chosen for further analysis to study the survivorship of this uncemented cup design in comparison to cemented references.Since 1980, The Finnish Arthroplasty Register has collected information on THAs on a national level. This study was based on information of THAs recorded in the Finnish Arthroplasty Register from 1985 to 2013. If patient had undergone bilateral THA, only the first one was included. Altogether, 49,289 THAs were included.The overall 10-year implant survivorship of uncemented implants was 84.3% (95% CI 83.5-85.0%) and of cemented implants 90.6% (90.2-90.0%). The 10-year survival of the contemporary uncemented implants was 90.8% (94.0-95.5%) and 88.5% (84.2-91.7%). The most recently introduced uncemented cup designs had good survivorship rates at 5 years; 10-year survival data are not yet available.We found that modern uncemented cup designs may provide good long-term survivorship rates comparable with the gold standard of cemented cups. However, polyethylene wear, liner problems and periosthetic osteolysis may still occur, also with modern uncemented acetabular components.
Total hip arthroplasty (THA) with uncemented implants has been a relatively common procedure in Finland during the recent decades. The most common brand of uncemented implants was chosen for further analysis to study the survivorship of this uncemented cup design in comparison to cemented references.Since 1980, The Finnish Arthroplasty Register has collected information on THAs on a national level. This study was based on information of THAs recorded in the Finnish Arthroplasty Register from 1985 to 2013. If patient had undergone bilateral THA, only the first one was included. Altogether, 49,289 THAs were included.The overall 10-year implant survivorship of uncemented implants was 84.3% (95% CI 83.5-85.0%) and of cemented implants 90.6% (90.2-90.0%). The 10-year survival of the contemporary uncemented implants was 90.8% (94.0-95.5%) and 88.5% (84.2-91.7%). The most recently introduced uncemented cup designs had good survivorship rates at 5 years; 10-year survival data are not yet available.We found that modern uncemented cup designs may provide good long-term survivorship rates comparable with the gold standard of cemented cups. However, polyethylene wear, liner problems and periosthetic osteolysis may still occur, also with modern uncemented acetabular components.