Evaluation of motion-correction methods for dual-gated cardiac positron emission tomography/computed tomography imaging




AuthorsRiku Klén, Tommi Noponen, Juha Koikkalainen, Jyrki Lötjönen,
Kris Thielemans, Erika Hoppela, Hannu Sipilä, Mika Teräs, Juhani Knuuti

PublisherLIPPINCOTT WILLIAMS & WILKINS

Publication year2016

JournalNuclear Medicine Communications

Journal name in sourceNUCLEAR MEDICINE COMMUNICATIONS

Journal acronymNUCL MED COMMUN

Volume37

Issue9

First page 956

Last page968

Number of pages13

ISSN0143-3636

DOIhttps://doi.org/10.1097/MNM.0000000000000539


Abstract
BackgroundDual gating is a method of dividing the data of a cardiac PET scan into smaller bins according to the respiratory motion and the ECG of the patient. It reduces the undesirable motion artefacts in images, but produces several images for interpretation and decreases the quality of single images. By using motion-correction techniques, the motion artefacts in the dual-gated images can be corrected and the images can be combined into a single motion-free image with good statistics.AimThe aim of the present study is to develop and evaluate motion-correction methods for cardiac PET studies. We have developed and compared two different methods: computed tomography (CT)/PET-based and CT-only methods.MethodsThe methods were implemented and tested with a cardiac phantom and three patient datasets. In both methods, anatomical information of CT images is used to create models for the cardiac motion.ResultsIn the patient study, the CT-only method reduced motion (measured as the centre of mass of the myocardium) on average 43%, increased the contrast-to-noise ratio on average 6.0% and reduced the target size on average 10%. Slightly better figures (51, 6.9 and 28%) were obtained with the CT/PET-based method. Even better results were obtained in the phantom study for both the CT-only method (57, 68 and 43%) and the CT/PET-based method (61, 74 and 52%).ConclusionWe conclude that using anatomical information of CT for motion correction of cardiac PET images, both respiratory and pulsatile motions can be corrected with good accuracy.



Last updated on 2024-26-11 at 20:08