A1 Vertaisarvioitu alkuperäisartikkeli tieteellisessä lehdessä

Measurement of brown adipose tissue mass using a novel dual-echo magnetic resonance imaging approach: A validation study




TekijätHolstila M, Virtanen KA, Gronroos TJ, Laine J, Lepomaki V, Saunavaara J, Lisinen I, Komu M, Hannukainen LC, Nuutila P, Parkkola R, Borra RJH

KustantajaW B SAUNDERS CO-ELSEVIER INC

Julkaisuvuosi2013

JournalMetabolism

Tietokannassa oleva lehden nimiMETABOLISM-CLINICAL AND EXPERIMENTAL

Lehden akronyymiMETABOLISM

Numero sarjassa8

Vuosikerta62

Numero8

Aloitussivu1189

Lopetussivu1198

Sivujen määrä10

ISSN0026-0495

DOIhttps://doi.org/10.1016/j.metabol.2013.03.002

Rinnakkaistallenteen osoitehttps://research.utu.fi/converis/portal/detail/Publication/1533590


Tiivistelmä
Objective
The aim of this study was to evaluate and validate magnetic resonance imaging (MRI) for the visualization and quantification of brown adipose tissue (BAT) in vivo in a rat model. We hypothesized that, based on differences in tissue water and lipid content, MRI could reliably differentiate between BAT and white adipose tissue (WAT) and could therefore be a possible alternative for 18 F-Fluorodeoxyglucose Positron Emission Tomography (18FDG-PET), the current gold standard for non-invasive BAT quantification.
Materials/Methods
Eleven rats were studied using both 18FDG-PET/CT and MRI (1.5 T). A dual echo (in-and-out-of-phase) sequence was used, both with and without spectral presaturation inversion recovery (SPIR) fat suppression (DUAL-SPIR) to visualize BAT, after which all BAT was surgically excised. The BAT volume measurements obtained via 18FDG-PET/CT and DUAL-SPIR MR were quantitatively compared with the histological findings. All study protocols were reviewed and approved by the local ethics committee.
Results
The BAT mass measurements that were obtained using DUAL-SPIR MR subtraction images correlated better with the histological findings (P = 0.017, R = 0.89) than did the measurements obtained using 18FDG-PET/CT (P = 0.78, R = 0.15), regardless of the BAT metabolic activation state. Additionally, the basic feasibility of the DUAL-SPIR method was demonstrated in three human pilot subjects.
Conclusions
This study demonstrates the potential for MRI to reliably detect and quantify BAT in vivo. MRI can provide information beyond that provided by 18FDG-PET imaging, and its ability to detect BAT is independent of its metabolic activation state. Additionally, MRI is a low-cost alternative that does not require radiation.

Ladattava julkaisu

This is an electronic reprint of the original article.
This reprint may differ from the original in pagination and typographic detail. Please cite the original version.





Last updated on 2024-26-11 at 22:13