Refereed journal article or data article (A1)

Midterm risk of cancer with metal-on-metal hip replacements not increased in a Finnish population




List of AuthorsElina Ekman, Inari Laaksonen, Antti Eskelinen, Pekka Pulkkinen, Eero Pukkala, Keijo Mäkelä

PublisherTaylor and Francis Ltd

Publication year2018

JournalActa Orthopaedica

Journal name in sourceActa Orthopaedica

Volume number89

Issue number9

Start page575

End page579

Number of pages5

ISSN1745-3682

eISSN1745-3674

DOIhttp://dx.doi.org/10.1080/17453674.2018.1487202

Self-archived copy’s web addresshttps://research.utu.fi/converis/portal/detail/Publication/35433055


Abstract

Background
and purpose — Metal-on-metal (MoM) total hip arthroplasty (THA) and hip
resurfacing arthroplasty (HRA) have been widely used during the early
21st century. We assessed the midterm risk of cancer of patients treated
with modern MoM hip implants compared with patients with non-MoM hip
implants and the general Finnish population with special interest in
soft tissue sarcomas and basalioma due to the findings of our previous
report.

Patients
and methods — All large-diameter head MoM THAs and hip resurfacings
performed in Finland between 2001 and 2010 were extracted from the
Finnish Arthroplasty Register (10,728 patients). Patients who underwent
conventional THA formed the non-MoM reference cohort (18,235 patients).
Data on cancer cases up to 2014 were extracted from the Finnish Cancer
Registry. The relative risk of cancer in the general population was
expressed as the ratio of observed to expected number of cases, i.e.,
standardized incidence ratio (SIR). Poisson regression analysis was used
to compare the cancer risk between the cohorts. The mean follow-up was
7.4 years (1–14) in the MoM cohort and 8.4 years (1–14) in the non-MoM
cohort.

Results
— The overall risk of cancer in the MoM cohort was comparable to the
general Finnish population (SIR 0.9, 95% CI 0.9–1.0). Risk of basalioma
in the MoM cohort was higher than in the general Finnish population (SIR
1.2, CI 1.1–1.4) and higher than in the non-MoM cohort in the
stratified regression analysis (RR 1.2, CI 1.0–1.4, p = 0.02). The SIR
of soft-tissue sarcoma in the MoM cohort was 1.4 (CI 0.6–2.8); the
incidence was same as in the non-MoM cohort.

Interpretation
— Metal-on-metal hip implants are not associated with an increased
overall risk of cancer during midterm follow-up.