A4 Article in conference proceedings
A miniaturized low power biomedical sensor node for clinical research and long term monitoring of cardiovascular signals




List of Authors: Jarno Tuominen, Eero Lehtonen, Mojtaba Jafari Tadi, Juho Koskinen, Mikko Pänkäälä, Tero Koivisto
Publication year: 2017
Book title *: 2017 IEEE International Symposium on Circuits and Systems (ISCAS)
ISBN: 978-1-5090-1427-9
eISBN: 978-1-4673-6853-7
ISSN: 2379-447X

Abstract

We present a low power, miniaturized biomedical sensor node featuring electro-, seismo-and gyrocardiography. Measurement session can be set up and controlled over Bluetooth Low Energy interface. Acquired raw or preprocessed sensor data can be stored locally to the microSD memory card. To the best of the authors' knowledge, the presented device is the smallest reported to enable heart monitoring from the electrical and mechanical (with six degrees of freedom) viewpoint. The average power consumption of the device is less than 15mW, enabling continuous acquisition of high quality data of more than 60 hours with the used 250mAh Lithium-Polymer battery.


Downloadable publication

This is an electronic reprint of the original article.
This reprint may differ from the original in pagination and typographic detail. Please cite the original version.




Last updated on 2019-29-01 at 22:44